3nh TS8560 benchtop spectrophotometer with dual LED and xenon sources and a rotatable screen
Spectrophotometers

TS8560 Benchtop Spectrophotometer (LED + xenon)

A flagship benchtop spectrophotometer with dual LED + xenon light sources, reflectance and transmittance (haze) measurement and a Φ154 mm sphere — with a rotatable 7" screen, 32 GB storage and automatic temperature/humidity compensation.

  • Dual light source LED + xenon lamp — better identification of fluorescent samples
  • Reflectance (SCI/SCE) and transmittance plus haze (per ASTM D1003) in a single instrument
  • Four aperture ranges (XLAV/LAV/MAV/SAV) and a transmission chamber for samples up to 54 mm thick
  • Rotatable 7" touchscreen, 32 GB storage, camera positioning, temperature/humidity compensation

The TS8560 is a 3nh flagship benchtop laboratory spectrophotometer with a dual light source — a full-spectrum LED plus a xenon lamp — giving it markedly better fluorescent-sample recognition than standard sphere spectrophotometers. It measures both reflectance (D/8 geometry, SCI/SCE, UV on/off) and transmittance plus haze (per ASTM D1003), covering opaque, transparent and liquid samples with one unit.

A Φ154 mm integrating sphere and four aperture ranges (XLAV/LAV/MAV/SAV) let you match the reading to the sample size, while the open transmission chamber accepts samples up to 54 mm thick. A rotatable 7" touchscreen, 32 GB storage, camera positioning and automatic temperature/humidity compensation make it a convenient, stable laboratory hub.

Key specifications

Parameter Value
Measurement geometry Reflectance D/8 (SCI/SCE, UV on/off); Transmittance D/0; haze (ASTM D1003); compliant with CIE No. 15, ISO 7724-1, ASTM E1164, DIN 5033 Teil 7, JIS Z8722
Integrating sphere Φ154 mm
Light source combined LEDs (with UV, 400/420 nm filters) and xenon lamp, 360–780 nm
Monochromator / sensor concave grating / dual CMOS sensor (256-element)
Range / resolution 360–780 nm, 10 nm interval, 5 nm half-bandwidth
Measuring aperture reflective: XLAV Φ25.4/30, LAV Φ15/18, MAV Φ8/10, SAV Φ4/6 mm; transmissive: Φ25.4 mm (up to 54 mm thick)
Colour spaces CIE L*a*b*, XYZ, Yxy, LCh, CIE LUV, Munsell, s-RGB, Hunter Lab, βxy, DIN Lab99
Colour-difference formulas ΔE*ab, ΔE*uv, ΔE*94, ΔE*cmc(2:1), ΔE*cmc(1:1), ΔE*00, DIN ΔE99, ΔE(Hunter)
Additional functions fluorescent-sample measurement, haze, WI/YI indices, temperature/humidity compensation
Display / memory rotatable 7" TFT touchscreen / 32 GB internal storage
Positioning camera preview, measurement in multiple orientations (side, upward, downward)
Interface / power USB, Bluetooth, printer port / DC 24 V adapter

Applications

The TS8560 targets the most demanding laboratory colour control — in paint, ink, textile and plastics production, printing, fluorescent samples, and anywhere simultaneous, accurate reflectance, transmittance and haze measurement is required.

Need help matching an instrument to your application? Request a quote — we will help you choose the right configuration and accessories.

Datasheet (PDF) — coming soon

Ask about this product

Looking for the right product?

Get in touch — we'll help you find the right solution.

Contact us